98
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND ARGUMENT 1 ARGUMENT 2
ARGUMENT 3
ARGUMENT 4 RESPONSE
R
Spec. Gravity G as ratio
of molecular weight.
(no units)
<Value>
(0.001-999.0)
OGR:<Value>
Example: OGR:0.105
No Argument
(ret. cur. set.)
OGR:<Value>
Example: OGR:0.105
C
Critical Temperature
and Pressure
<Value> Crit. Temp.
(Deg. F) [-459–99999]
<Value> Crit. Pressure
(PSIA) [0-99999]
OGC:<Value>,<Value>
Example: OGC:50.0,100.0
No Argument
(ret. cur. set.)
OGC:<Value>,<Value>
Example: OGC:50.0,100.0
Z
Compressibility factor
at flowing TP (applies
only for T/P meters)
<Value> [0.0 – 1.0] OGZ:<Value>
Example: OGZ:0.325
No Argument
(ret. cur. set.)
OGZ:<Value>
Example: OGZ:0.325
W
Molecular weight (Lb/
LbMol) (applies only for
Ratio of Densities mode)
<Value>
[0.0 – 9999.9]
OGW:<Value>
Example: OGW:18.3581
No Argument
(ret. cur. set.)
OGW:<Value>
Example: OGW:18.3581
Flowing
Units
Parameters
Sets/Reads Temperature
and Pressure parameters
for STD, NORMAL,
and Customer Actual
units.
NOTE: Actual Units
parameters equal to the
Actual Flowing Conditions.
Customer Actual Units
Temperature and Pressure
can be used to change
flow conditions for
meters without T/P option.
29 FP S
Standard Units
Temperature and
Pressure
<Value> STD Temp.
(Deg. F) [-459–99999]
<Value> STD Pressure
(PSIA) [0-99999]
FPS:<Value>,<Value>
Example: FPS:70.0,14.7
No Argument
(ret. cur. set.)
FPS:<Value>,<Value>
Example: FPS:70.0,14.7
N
Normal Units
Temperature and
Pressure
<Value> NORM Temp.
(Deg. C) [-273–99999]
<Value> NORM Press
(KPa) [0-99999]
FPN:<Value>,<Value>
Example: FPN:0.0,101.3
No Argument
(ret. cur. set.)
FPN:<Value>,<Value>
Example: FPN:0.0,101.3
C
Customer Actual Units
Flowing Temperature
and Pressure (see
Note in description)
<Value> Actual Temp.
(Deg. F) [-459–99999]
<Value> Act. Pressure
(PSIA) [0-99999]
FPC:<Value>,<Value>
Example: FPC:50.0,100.2
No Argument
(ret. cur. set.)
FPC:<Value>,<Value>
Example:FPC:50.0,100.2